=== Topic: Cross-calibration ===  | |||
* Following round-robin reference samples have been used so far for SESANS at several sources (according to Wim Bouwman): Flexi-Por, Papyex. These measurements have also been compared to SANS/USANS results from the same samples.  | |||
* But: these kinds of samples have the disadvantage that they show artifacts when used for TOF-SESANS.  | |||
* Also, the presented reference data had differences in peak-minima of measured reverence samples at different sources. The reason for this difference still has to be identified.  | |||
* General consens: We need more sets of standard samples at all facilities  | |||
=== Notes from the discussion round: ===  | |||
* What is needed: New standard samples with shorter length scales of around 50nm (but such gratings are difficult to make)  | |||
* The question was raised if new round robin samples should be fabricated which can be handed from facility to facility. On the other hand, a good reference sample depends on each use-case - there is no perfect standard sample for several techniques/facilities. Instead, good existing standard samples (like the above mentioned and already measured Flexi-Por/Papyex) should be used.  | |||
Latest revision as of 13:58, 12 March 2025
Notes from canSAS 2024 Second Working Group Discussion Session
Includes:
- Data Formats
 - TOF SANS
 - SESANS
 - Tomograph
 
SESANS Session Notes SAS2024 - Sun 2024-11-03
Topic: Cross-calibration
- Following round-robin reference samples have been used so far for SESANS at several sources (according to Wim Bouwman): Flexi-Por, Papyex. These measurements have also been compared to SANS/USANS results from the same samples.
 - But: these kinds of samples have the disadvantage that they show artifacts when used for TOF-SESANS.
 - Also, the presented reference data had differences in peak-minima of measured reverence samples at different sources. The reason for this difference still has to be identified.
 - General consens: We need more sets of standard samples at all facilities
 
Notes from the discussion round:
- What is needed: New standard samples with shorter length scales of around 50nm (but such gratings are difficult to make)
 - The question was raised if new round robin samples should be fabricated which can be handed from facility to facility. On the other hand, a good reference sample depends on each use-case - there is no perfect standard sample for several techniques/facilities. Instead, good existing standard samples (like the above mentioned and already measured Flexi-Por/Papyex) should be used.